BACKGROUND
Reactive atrial-based antitachycardia pacing (rATP) aims to terminate atrial tachyarrhythmia/atrial fibrillation (AT/AF) episodes when they spontaneously organize to atrial flutter or atrial tachycardia; however, its effectiveness in the real-world has not been studied. We used a large device database (Medtronic CareLink, Medtronic, Minneapolis, MN, USA) to evaluate the effects of rATP at reducing AT/AF.
METHODS
Pacemaker, defibrillator, and resynchronization device transmission data were analyzed. Eligible patients had device detected AT/AF during a baseline period but were not in persistent AT/AF immediately preceding first transmission. Note that 1:1 individual matching between groups was conducted using age, sex, device type, pacing mode, AT/AF, and percent ventricular pacing at baseline. Risks of AT/AF events were compared between patients with rATP-enabled versus control patients with rATP-disabled or not available in the device. For matched patients, AT/AF event rates at 2 years were estimated by Kaplan-Meier method, and hazard ratios (HRs) were calculated by Cox proportional hazard models.
RESULTS
Of 43,440 qualifying patients, 4,203 had rATP on. Matching resulted in 4,016 pairs, totaling 8,032 patients for analysis. The rATP group experienced significantly lower risks of AT/AF events lasting ≥1 day (HR 0.81), ≥7 days (HR 0.64), and ≥30 days (HR 0.56) compared to control (P < 0.0001 for all). In subgroup analysis, rATP was associated with reduced risks of AT/AF events across age, sex, device type, baseline AT/AF, and preventive atrial pacing.
CONCLUSIONS
Among real-world patients from a large device database, rATP therapy was significantly associated with a reduced risk of AT/AF. This association was independent of whether the patient had a pacemaker, defibrillator, or resynchronization device.